Important Information — Please Read Before Use
9
ENF-V3 OPERATION MANUAL
WARNING
• Do not use high-frequency electrosurgical equipment with this endoscope. The
electrode of the electrosurgical equipment can get in contact with the endoscope,
and it may cause patient injury, burns, bleeding, perforations, and/or equipment
damage.
• Do not use the laser unit with this endoscope. Endoscopic observation may not be
performed normally.
CAUTION
• Do not pull the universal cord and video cable during an examination. The light
guide connector will be pulled out from the output socket of the light source and the
endoscopic image will not be visible.
• Do not coil the insertion tube, universal cord, or video cable with a diameter of less
than 10 cm. Equipment damage can result.
• Do not attempt to bend the endoscope’s insertion section with excessive force. The
insertion section may be damaged.
• Do not apply shock to the distal end of the insertion section, particularly the
objective lens surface at the distal end. Visual abnormalities may result.
• When the endoscope is dropped or the distal end of the endoscope receives a hard
impact, the endoscope may be damaged even if no visible damage of the lens on
the distal end can be found. In this case, stop using the endoscope, and contact
Olympus.
• Do not twist or bend the bending section with your hands. The endoscope damage
may result.
• Do not squeeze the bending section forcefully. The covering of the bending section
may stretch or break and cause water leaks.
• Do not put or press the video connector and light guide connector on the insertion
section when transporting or reprocessing. The insertion section may be damaged.
• Turn the video system center ON only when the video connector is connected to
the video system center. In particular, confirm that the video system center is OFF
before connecting or disconnecting the video connector. Failure to do so can result
in equipment damage, including destruction of the image sensor.
• The endoscope’s remote switches cannot be removed from the control section.
Pressing, pulling, or twisting them with excessive force can break the switches
and/or cause water leaks.
• Do not hit or bend the electrical contacts on the video connector. The connection to
the video system center may be impaired and faulty contact can result.
• Do not pull the video cable during an examination. The endoscopic image may not
be visible.