3.6 Inspection of the endoscopic system
33
ENF-V3 OPERATION MANUAL
Ch.3
Inspection of the remote switches
WARNING
All remote switches should be checked to work normally even if they are not
expected to be used. The endoscopic image may freeze, or other irregularities may
occur during examination and may cause patient injury, bleeding, and/or
perforation.
Figure 3.15
6
Turn the UP/DOWN angulation control lever
slowly in each direction until it stops.
Figure 3.14
7
Confirm that the WLI and NBI endoscopic images do not momentarily disappear or
display any other irregularities.
1
Depress every remote switch.
2
Confirm that the specified functions work normally.
Remote switches