Rotatable Clip Fixing Device
41
Chapter 4 Emergency Treatment
Chapter 4 Emergency Treatment
4.1
Emergency Treatment
Do not try to forcibly withdraw the
instrument from the endoscope if the clip
cannot be detached from the instrument.
Forcibly withdrawing the instrument
could cause patient injury such as
perforation, bleeding or mucous
membrane damage.
If the clip cannot be detached from the instrument, follow
the procedures described in this section.
If the distal end of the coil sheath or clip pipe is crushed
or deformed, it may not be possible to detach the clip
from the instrument.