Rotatable Clip Fixing Device
39
Chapter 3 Preparation, Inspection and Operation
Removal of an Aspirated Clip or Clip Connector
If a clip or clip connector is accidentally aspirated into the
endoscope, follow the procedure below to remove it.
1.
Withdraw the endoscope from the body cavity,
keeping the insertion portion and bending section
straight. Leave the biopsy valve mounted on the
endoscope.
2.
Remove the suction tube and connect a syringe
filled with tap water to the endoscope’s suction
connector (see Figure 3.16).
Figure 3.16
3.
While gently pressing the suction valve, inject the
tap water into the endoscope’s suction connector
(see Figure 3.17).