Chapter 3 Preparation, Inspection and Operation
Rotatable Clip Fixing Device
26
Attaching the Clip
• Do not use a clip that has not been
properly attached. Otherwise, the clip
may not operate correctly or it could be
damaged.
• After attaching the clip, dispose of the
cartridge properly. Otherwise, infection
may result.
•
After retracting the clip into the coil
sheath, do not extend the clip out from
the tube sheath until the clip is actually
going to be used. If the clip is extended
unnecessarily, it may be impossible to
insert it into the endoscope.
• Do not disassemble a used cartridge and
attempt to reuse the clip in it. Otherwise,
the clip cannot be attached properly or
may be damaged.
To attach additional clips after the first
one, detach the clip connector from the
hook by following the instructions in
“Detaching the Clip Connector” on
page 38.
If the clip cannot be attached securely, do not use the
clip and/or instrument; use a spare instead.
1.
Open the package containing the cartridge.
2.
Carefully pull the slider up to the ring (yellow).