Rotatable Clip Fixing Device
19
Chapter 3 Preparation, Inspection and Operation
3.2
Inspection
Wear the personal protective equipment as specified in
the table on page 50.
Before each case, always inspect the instrument and clip
according to the following procedures.
If an abnormality in the instrument or clip is detected, use
a spare instrument or clip, inspecting it thoroughly before
use.
Inspection of the Sterile Package
The clip is a single-use, disposable item.
Do not use or attempt to sterilize it.
Reusing the clip could pose an infection
control risk, cause tissue irritation or
malfunction.
Inspect the sterile package for tears, inadequate sealing
or water damage. If the sterile package shows any
irregularities, the sterile condition of the instrument or clip
may have been compromised. Use a spare instead.