Rotatable Clip Fixing Device
23
Chapter 3 Preparation, Inspection and Operation
Inspection of Operation
If the instrument does not operate smoothly and as
intended, do not use the instrument; use a spare instead.
1.
Holding the instrument as shown in Figure 3.4, form
a loop in the insertion portion approximately 20 cm
in diameter.
Figure 3.4
2.
Operate the slider and confirm that the hook retracts
into and extends from the coil sheath smoothly and
as intended.
Approximately 20 cm.
Extend
Retract