Lamp life meter
reset function
check
Panel operation
check
The lamp life meter does not return to zero
even if the lamp life meter reset switch.
Does LED indicate
zero when the lamp life
meter reset switch is
depressed?
N o
Yes
N o
Is the reset switch held
depressed?
TROUBLESHOOTING CLV-160
Yes
Hold the reset switch
depressed?
N o
Does LED indicate
zero?
Yes
Replace the UPCL16CR00.
(Replace the main board.)
End