TROUBLESHOOTING CLV-160
Filter switching
The turret plate does not rotate even
operation check
if the filter switch is depressed.
Does the turret
return to the initial
position when the power
is turned on?
No
Yes
Check the harness
connection.
UPCL16CR00
Stepping motor
UPCL16CR00
Front panel U
Check that the turret
plate is not extremely
loose.
Yes
Does the turret
return to the
initial position?
No
Replace the limit switch.
Yes
Does the turret
return to the
initial position?
No
Replace the
UPCL16CR00. (Replace
the main board.)
Yes
Does the turret
return to the
initial position?
No
Replace the turret U.