Instructions for Use
EN
Celon
Lab
ENT
W7098046
11
Audio-frequency if impedance too high
≈ 2.5 kHz, pulsed with approx. 1 Hz
Rated supply voltage
100 … 120 V~
200 ... 240 V~
Rated supply frequency
50 … 60 Hz
Power input at max. power
95 VA
Power fuses (2 x)
1.6 A T-lag, 250 V, 5x20 mm, UL/CSA
REF: WB991007 (100 ... 120 V~)
0.8 A T-lag, 250 V, 5x20 mm, UL/CSA
REF: WB991001 (200 ... 240 V~)
Power connection line
IEC 60320-1/C13
Dimensions (W x H x D)
267 x 165 x 335 mm
Weight
Unit:
7.3 kg
Footswitch:
1.0 kg
Protection class
I
Classification according to Medical Devices
Directive:
Europe / US (FDA)
IIb / II
Foreign body/moisture protection
IP 21, footswitch IP 68
Transport conditions:
Temperature
Relative humidity
-34 ... +65 °C
15 ... 80 %, non-condensing
Storage conditions:
Temperature
Relative humidity
-25 ... +60 °C
15 ... 80 %, non-condensing
Operating conditions:
Temperature
Relative humidity
+10 ... +40 °C
15 ... 80 %, non-condensing
Standards and directives applied:
93/42/EEC MDD
Conformity assessment
EN ISO 13485
Quality assurance systems – Medical Devices –
Specific requirements for application as per ISO 9001
EN 60601-1
Medical electrical devices
EN 60601-1-2
Medical electrical devices (EMC)
EN 60601-2-2
Medical electrical devices (electrosurgical devices)
EN ISO 14971
Medical devices risk analysis