10
W7098046
6.2 Repairs
If a defect occurs, Olympus Winter & Ibe GmbH or a contact appointed by
Olympus Winter & Ibe GmbH (e.g. distributor) must be informed. The repair
must only be carried out by Olympus Winter & Ibe GmbH or a firm authorised by
Olympus Winter & Ibe GmbH (ask the Olympus Winter & Ibe GmbH distributor
responsible for your area). Service documents such as circuit diagrams, parts
lists, equipment descriptions and setting instructions are available from the
manufacturer for technicians who are authorised to carry out maintenance and
repair.
Caution:
When the housing is opened there is a danger of electric shock. The
unit must only be opened by authorised technicians.
6.3 Safety Checks
The unit must undergo a safety check at yearly intervals in accordance with the
national statutory regulations. For this purpose, Olympus Winter & Ibe GmbH or
a firm authorised by Olympus Winter & Ibe GmbH must be commissioned by the
operator.
6.4 Disposal of the Unit
When disposing of the product, or any of its components, follow all applicable
national and local laws and guidelines.
Waste Electrical and Electronic Equipment
In accordance with European Directive 2002/96/EC on Waste Electrical
and Electronic Equipment, this symbol indicates that the product must
not be disposed of as unsorted municipal waste, but should be
collected separately.
Refer to your local Olympus distributor for return and/or collection systems
available in your country.
7. Technical Data
Unit designation
CelonLab
ENT
power control unit
REF: WB991007 (100 ... 120 V~)
REF: WB991001 (200 ... 240 V~)
Outputs
bipolar
Applied part
Type BF, defibrillation-safe
Rated output power
25 W at 100 Ohm
Power selection
From 1 to 25 W in 1 W steps
Impedance dependency of the power
See information accompanying the unit
Rated output frequency
470 kHz
Max. peak output voltage
< 150 V
peak
Max. peak output voltage as a function
of the power level
See information accompanying the unit
Audio-frequency of “generator active”
signal
300 Hz – 2.5 kHz (for impedance range
25 – 400 ohms)
Audio-frequency if impedance too low
300 Hz, pulsed with approx. 1 Hz