25
Using the Joint Plates
The joint plates @ can be attached and locked between the slider knob
and slider as shown in the figure. The joint plates should be attached on
both ends of the filter slider.
By locking with the joint plates, you can switch the barrier and excitation
filters together as a set.
Fig. 6
NOTES
· When inserting the 6-position filter slider in the 6-position filter slider near
the rear panel, insert from the left so that the “U-RSL6” inscription comes
at the deep. Otherwise, the filters will not be set in the correct positioning.
· When the 6-position filter slider near the rear panel is used, avoid using
the interference type or color glass type filters. This is because the 6-
position filter slider near the rear panel is one of the positions where the
energy from the light source is concentrated. When an interference type
or color glass type filter is mounted in it, the filter interference film may
peel off or the color glass may be damaged.
· Make sure that the 6-position filter slider is set to a click position.
· For safety, insert the provided light shield plates in the unused filter posi-
tions.
Fig. 7
@