23
Item
Specification
SPECIFICATIONS
Vertical illuminators
Reflected Illuminator
BX-URA2
Fluorescence Illuminator
BX-RFA
· UIS2/UIS (Universal Infinity System) optical system (featuring infinity correction)
· Magnification: 1X (Superwide field: NA 26.5)
· Observation switching: Mirror unit turret carrying max. 6 mirror units.
· Aperture iris diaphragm and field iris diaphragm (Both centerable) Detachable with the BX-
RFA.
· Shutter provided.
· Slider inlet
@ Analyzer/6-position barrier filter slider
² Polarizer/6-position filter slider
³ ND filters
@ Analyzer/6-position barrier filter slider
² 6-position filter slider
³ ND filters
| 6-position filter slider
· Available observation modes
@ Reflected fluorescence
² Reflected fluore Transmitted DIC
³ Reflected fluore Phase contrast
| Reflected light brightfield
ƒ Reflected light darkfield
… Reflected light DIC
† Reflected light simplified polarization
‡ Transmitted light
@ Reflected fluorescence
² Reflected fluore Transmitted DIC
³ Reflected fluore Phase contrast
| Transmitted light
· Optional accessories
@ Exciter/balancer
² Rectangle field stop
Mercury lamp housing
· 100 W mercury lamp housing U-LH100HG
· 100 W mercury apo lamp housing U-LH100HGAPO
· Mercury burner: USH-103OL (USHIO) or HBO103W/2 (OSRAM)
Operating environment
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104° F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F), decreasing linearly
through 70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations; Not to exceed ±10% of the normal voltage.
· Pollution degree: 2 (in accordance with IEC60664)
· Installation/Overvoltage category: II (in accordance with IEC60664)