NOTE
Before connecting to the power supply, check that the supply voltage
agrees with the electrical data on the type plate of the peristaltic pump
unit.
If the voltage of the facility is different from the voltage indicated on the
type plate of the peristaltic pump unit, contact Olympus or your
distributor.
6.4 Equipotential bonding
To increase electrical safety, the unit can be connected to the equipotential bonding
system of the treatment room. All equipment housings that come into contact with the
patient are electrically connected in order to prevent low-frequency electrical currents
from endangering the patient in the event of a defect in the conventional protective
conductor system.
6.5 Connection of footswitch
Before connecting, confirm that the
footswitch plug (22)
is free of scratches and cracks
and that the
footswitch pedal (21)
is not damaged. Align the
footswitch plug (22)
so that
the keying is facing down. Insert the
footswitch plug (22)
into the
footswitch connector
(15)
of the peristaltic pump unit on the rear panel and rotate the fastener ring fully
clockwise to tighten it (see fig. 6.3).
Fig. 6.3: Connection of the footswitch
CAUTION
The footswitch plug is not waterproof, and liquid such as water must not
get into the plug.
Do not connect products other than the footswitch for this peristaltic
pump unit to the
footswitch connector (15)
. Otherwise, the footswitch
might not function and may cause patient injury and / or damage of the
equipment.
Fastener ring