CERO_Operators Guide_Version 2018-12_01
59
8.2 CO
2
Level
8.2.1 CO
2
Alarm Triggered
Possible explanation
Recommended action
1. Malfuntioning / empty CO
2
supply
Check the integrity of the CO
2
supply and its
connection to the CERO
2. CERO not properly sealed
Check all cover are properly tight, convection
channel in place
3. CO
2
Level too high
See chapter 8.2.2
4. CO
2
Level too low
See chapter 8.2.3
5. Unidentified
Switch off the CO
2
supply and call customer support
8.2.2 CO
2
Level Too High
Possible explanation
Recommended action
1. Offset value too high
Readjust CO
2
offset (chapter 5.5)
2. Convection channel defect
Replace convection channel (chapter 7.4)
3. Input pressure too high
Reduce input pressure to 1.0 ±0.2 bar (14.5 ±3.0 psi)
4. Filter to sensing unit overly dirty
or partially plugged
Replace HEPA filter (chapter 7.5)
5. CO
2
sensor incorrectly calibrated
Call customer support
6. CO
2
sensor failure
Call customer support
7. CO
2
valve defect (no clack audible) Call customer support