CERO_Operators Guide_Version 2018-12_01
54
ATTENTION
Spraying directly on the unprotected reader plugs may cause damage and
may cause an electric short of the device.
5.
Spray the CERO Culture Chamber with MICROLAB
®
disinfectant spray and wipe using a lint
free cloth.
6.
Replace the 3 Hepa Filters.
7.
If the Tube Holders are heavily soiled you may remove them (see Paragraph 7.6) and soak
them in MICROLAB
®
Disinfectant or replace them.
8.
Replace the Convection Channel (see Paragraph 7.4).
7.3 Remove and Re-Install the Front Cover
1.
Open the Sliding Top Cover.
2.
Switch off the CERO.
3.
Loosen the thumb screw securing the front cover (see figure 45).
4.
Lift the Front Cover to dis-engage it from the 4 latch slots.
5.
Remove the Front Cover.
Um die Abdeckung wieder anzubringen, gehen Sie wie folgt vor:
1.
Place the Front Cover in back onto its latches.
2.
Press it down firmly.
3.
Fix the front thumb screw.
1. Loosen the thumb
screw fixing the front
cover
2. Lift front cover on the right and on the left to free it from
ist latches. Then pull the cover.
Figure 45: Removing the fron t cover