CERO_Operators Guide_Version 2018-12_01
26
4. Connect CO
2
supply
□
Connect the CERO to the low pressure side of a CO
2
regulator using the appropriate
CO
2
Tubing (Tygon® Tubing with 3/16" (4.8 mm) inner diameter (figure 14).
□
Make sure that the connection is tight and secure.
Figure 14: CO
2
supp ly conn ection
IMPORTANT NOTE
The proper and secure connection of the CO
2
supply to the CERO is the sole
responsibility of the user.
□
Set the output pressure of the regulator to 1.0 ± 0.2 bar (14.5 ± 3 psi).
5. Perform the decontamination procedure
□
See Chapter 7.2.