NOTE:
Use #18 AWG (0.9 mm²) shielded cable for remote sensors. Maximum
distance is 200 meters.
NOTE:
When wiring remote sensors to the
, “485 B” on J3 should be
connected to “B-” in the remote sensor enclosure, and “485 A” on J3 should
be connected to “A+” in the remote sensor enclosure.
NOTE:
For remote or standalone sensors, there are four terminal blocks
located in the remote sensor housing. These terminal blocks are all tied
together and follow the same wiring scheme mentioned above.