51
If you initiate the change span
option, the status bit will start to
flash and the
span value can
now be changed. The current span
value is displayed on the top line of
the LED display. Use the “↑” and “↓”
keys to change the span value.
When the desired value is reached,
press the “ ” key to save changes.
Pressing the “Mode” key or letting
the timer count down to zero without
saving the new value, will take you
back into the Programming Mode.
NOTE:
If the “ ” key is not pressed,
the new span value will not be
saved.
NOTE:
Span Gas Concentration for
combustibles can be set from 0% to
100%LEL. For the sake of resolution,
we
suggest
that
Span
Gas
Concentration should be set above
20% LEL.
Figure 5-25 Sample Span Gas
Concentration Display
Figure 5-26 Flashing Status Bit
Sensor Span Reserve
The
span
reserve
option
is
designated with an “r” in the status
bit. The current span reserve is
displayed on the top line of the LED
display.
NOTE:
There are a few other options
that appear that do not have any
function associated with them.
These are reserved for future
functionality.
Figure 5-27 Sample Span Reserve
Display
# # #
Содержание iTrans 2
Страница 6: ...6...
Страница 9: ...9...
Страница 10: ...10...
Страница 19: ...19 Figure 2 5 Electronics Board for Remote Sensor...
Страница 20: ...20...
Страница 22: ...22 Figure 3 1 Mounting the Gas Monitor on a Wall Figure 3 2 Mounting the Gas Monitor on a Column Using U Bolts...
Страница 28: ...28 Figure 4 4 Wiring Diagram for a Single On board Sensor...
Страница 31: ...31 Figure 4 7 Wiring Remote Sensors Back to...
Страница 52: ...52...
Страница 60: ...60...
Страница 66: ...66...
Страница 72: ...72 Figure A 4 Wiring Diagram of Single Sensor HART Supported...
Страница 73: ...73 Figure A 5 Wiring Diagram of Dual Sensor HART Supported...
Страница 80: ...80 Figure A 15 EDD GUI View Figure A 16 PC to HART Interface Wiring Diagram...
Страница 86: ...86...
Страница 90: ...90...
Страница 98: ...98...
Страница 100: ...100...
Страница 104: ...104...
Страница 105: ...105...