56
Addr
Inst
R/W
Host
R/W
Range
Description
$0107 = 1 decimal place for gas type HCN
$0002 = 0 decimal places for gas type H
2
S
$0206 = 2 decimal places for ClO
2
40105
W
R/W
$0000 to $FFFF
Instrument Mode
Holds code for current mode of
instrument. Possible working modes of
instrument are listed below.
$0001
Normal
$0002
Calibration
$0003
Warm-up
$0006
Zeroing
$0008
Fault
$0009
Reset
Examples:
Sensor in zero fault = $0008
Sensor zeroing
= $0006
40106
W
R
$0000 to $FFFF
Status Bits
Holds 16 bits of status for various
parameters in the instrument. A bit value
of “1” indicates that the associated fault
condition is present.
Bit 15
= current loop open
Bit 14
= current loop shorted
Bit 13
= power fault
Bit 12
= 5 volt fault
Bit 11
= missing sensor
Bit 10
=
(not defined)
Bit 6
= configuration fault
Bit 5
= zero fault
Bit 4
= calibration fault
Bit 3
= over-range
Bit 2
= failed sensor
Bit 1
= high alarm
Bit 0
= low alarm
Examples:
Missing sensor = Bit 11 is set
=
$0800
Power fault and
failed sensor = Bits 13 and 2 set
=
$2004
Содержание iTrans 2
Страница 6: ...6...
Страница 9: ...9...
Страница 10: ...10...
Страница 19: ...19 Figure 2 5 Electronics Board for Remote Sensor...
Страница 20: ...20...
Страница 22: ...22 Figure 3 1 Mounting the Gas Monitor on a Wall Figure 3 2 Mounting the Gas Monitor on a Column Using U Bolts...
Страница 28: ...28 Figure 4 4 Wiring Diagram for a Single On board Sensor...
Страница 31: ...31 Figure 4 7 Wiring Remote Sensors Back to...
Страница 52: ...52...
Страница 60: ...60...
Страница 66: ...66...
Страница 72: ...72 Figure A 4 Wiring Diagram of Single Sensor HART Supported...
Страница 73: ...73 Figure A 5 Wiring Diagram of Dual Sensor HART Supported...
Страница 80: ...80 Figure A 15 EDD GUI View Figure A 16 PC to HART Interface Wiring Diagram...
Страница 86: ...86...
Страница 90: ...90...
Страница 98: ...98...
Страница 100: ...100...
Страница 104: ...104...
Страница 105: ...105...