H301-MIZAR-TILT
User Manual
REV 02
www.oko-lab.com
Pa
ge
14
11.
Working with perfusion
H301-MIZAR-TILT features 4 perfusion holes on the sides of the chamber base for the insertion of perfusion tubing up
to 2.0 mm in outer diameter. Small screws plug the perfusion holes when not in use (Grub screws M2,5x3). Remove
the small screws as necessary before introducing perfusion tubing.
Figure 15 shows the location of the perfusion holes.
Openings for
Perfusion
1. Frontal view
2. 3D view
Figure 15. Perfusion
12.
Connection of the Chamber with Z stage
To fix the H301-MIZAR-TILT on the stage, follow the steps listed below and shown in Figure 16:
1.
Fix the bracket to the SLR.300 Piezoconcept Stage with four M2.5x6 screws (see Image 1 of Figure 16).
2.
Fix the base of the chamber with two M3x6 screws on the bracket (see Image 2 of Figure 16).
3.
Insert the sample holder (see paragraph 5, 6 or 7 according to the sample holder you have purchased).
1
2
Figure 16. Connection of the Chamber with Z Stage