8.
Working with 35mm petri-dish
Flat springs prevent movement of 35mm petri-dish inside the opening on the Chamber Base. To insert the MIZAR-1x35,
follow the steps shown in Figure 12 and listed below:
1.
Insert the MIZAR-1x35 into the Chamber Base (see image 1 in Figure 12).
2.
Place the magnetic clamps over the 1 MIZAR-1x35 (see image 2 in Figure 12), then push down the head of the
clamps simultaneously and move it in the direction shown in the image 3 of Figure 12.
3.
Turn the flat springs (see image 4 in Figure 12) and insert the petri-dish (see image 5 in Figure 12) into the sample
holder.
4.
Lock the petri-dish by using the flat springs (see image 6 in Figure 12).
5.
Close the Chamber Base with the Lid (see image 7 and 8 in Figure 12).