Chapter 1
7
Introduction
Controller Power Requirements
•
115/230 VAC, 50/60 Hz
•
Fuse rating: 115 VAC (±10%) - 1A Type T;
220/230/240 VAC (±10%) - 1/2A Type T
Hardware Requirements for Running WinPulse
•
Pentium/100 MHz IBM-compatible platform (minimum)
•
VGA monitor (or higher)
•
One serial port (RS-232-C) (16550 buffered UART 38.4 K baud required)
Host Software Requirements
•
Windows Version 3.1 or later, Windows 95, or Windows NT
Gas Requirements:
•
Carrier: helium, nitrogen, or hydrogen - 80 psig, 99.999% purity or better
•
Air: 60 psig, Zero Air 99.999% purity or better
•
Hydrogen: 60 psig, 99.999% or better (Electrolytic Grade)
Recommended: Use of hydrocarbon and oxygen filters for carrier gas.
Controller Board Outputs
•
2 Channels (0–1 V)
•
1 Serial RS-232-C
•
1 Signal In (electrometer; PFPD)
•
High Voltage Out (PMT 0–1000 V)
•
Ignitor Current (0–3.4 A)
•
Oscilloscope Output (20 Hz, 25 ms display)
•
GC “Start” Sense (senses contact closure)
•
S/W HV protection (PMT protection)
•
Timed events (from GC Start Sense): autozero, range, attenuation, ignitor, and
mode-channel (custom and record)