Chapter 3
27
Introduction to WinPulse
Event
Timed events that can be selected from a list by clicking on the
down arrow next to the Event field. Scroll down through the list
until the desired event is located and then click on it. Figure 3.4
shows an example of each event that can be selected. The selected
events will be initiated at the specified times.
Auto-Zero
Sets the zero offset on the selected channel.
Attenuation
Attenuates the output on the selected channel by the
specified value (values of 2
X
can be selected where x = 0
to 8). Note, by changing the attenuation, the signal to noise
ratio may change. (See Appendix D for a more detailed
discussion.)
Range
Sets the electrometer to the selected range (1, 10, or 100,
which correspond to 0.1, 1.0 and 10.0 nA/V, respectively).
Square Root
Enables or disables calculation of the square root of the
output signal in the selected channel. This supersedes the
Enable Sqrt selection set specified in the Gate Param-
eters screen. Enabling this option allows the operator to
linearize quadratic response signals associated with some
modes (e.g. sulfur mode, S).
Gate
Specifies which gate mode, listed in the Gate Parameters
List, is linked to the selected Channel mode (see below).
Record
Records the raw chemiluminescence data between the
specified start and stop times (in minutes) into a data file.
Ignitor
Sets the electric current delivered to the ignitor of the
PFPD.
Channel
Specifies the Channel that the Event occurs on.
Value
Sets the value that applies to the selected Event. The unit of
measure for the value varies by Event. For a recording event, the
value is the stop time.
The function of each button along the right side of the Timed Events screen is as
follows:
OK
Transmits the Event List to the Detector Controller and exits the
Timed Events screen.
Cancel
Disables all changes made to the timed events parameters and exits
the screen without downloading the changes to the Detector
Controller.
Help
Activates the Help screen.
Add
Appends a timed event to the Event List.
Содержание 5380
Страница 89: ...84 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 95: ...90 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 99: ...94 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 107: ...102 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 117: ...112 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 125: ...120 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 133: ...P O Box 9010 College Station TX 77842 9010 Tel 979 690 1711 FAX 979 690 0440 www oico com...