32
Model 5380 PFPD Operator’s Manual
Rev. 3.1
4. Carefully press the PFPD assembly and surrounding insulation into the
previously hollowed-out insulation plug in the selected detector port (see
Figure 4.3). Do not excessively bend the leads protruding from the heater/
sensor probes. Ensure that the detector base protrudes into the GC oven cavity
and that the detector mounting plate is flush with the metal cover in the
detector area.
5. Align the detector mounting plate with the screw holes corresponding to the
selected detector port and secure the mounting plate using the mounting
screws (Part #252585 for Agilent 6890 GC; Part #169963 for Agilent 5890
GC).
Note:
When mounted, the PMT protrudes to the right side of the GC. If this
orientation of the detector needs to be changed, refer to “Installing the
Combustor and Reassembling the PFPD” in this chapter.
Installing the Model 5380 PFPD in the
Agilent 6890 Valve Box
Following are instructions for installing the Model 5380 PFPD in the Agilent 6890
GC valve box area. The valve box sits on top of the Agilent 6890 and contains the
valves and plumbing, heated zones, sensors, and insulation. For more information,
see the Agilent 6890 Series Gas Chromatograph Operating Manual.
1. Cut out and remove the valve box plate. See Figure 4.4.
2. Remove the precut insulation layer.
Heater Cable
Figure 4.3. Placement of the PFPD and Attachment of Heater Cable
Model 5380 PFPD
Sensor Assembly
PMT HV and
PMT Signal Cables
Ingitor Cable
AIB Board
Signal Out Cable from
Detector Controller
Содержание 5380
Страница 89: ...84 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 95: ...90 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 99: ...94 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 107: ...102 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 117: ...112 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 125: ...120 Model 5380 PFPD Operator s Manual Rev 3 1 Notes...
Страница 133: ...P O Box 9010 College Station TX 77842 9010 Tel 979 690 1711 FAX 979 690 0440 www oico com...