![Ohkura RM10C Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/ohkura/rm10c/rm10c_instruction-manual_3527809056.webp)
―
55
―
HXPRM10mnC0001E
(7)
(Skip)
Setting SKIP on the Channel does not display and recording.
Example) When skipping Channel 6.
Hold down the
“
MENU
”
key for 3 seconds or more
to enter the setup mode. Use the
△
key to display
"
". Press the
“
ENT
”
key.
Use the
△
key to select the channel you want to set.
Press the
“
ENT
”
key.
Use the
△
key to select "
". Press the
“
ENT
”
key.
Press the
“
ENT
”
key. Setting is completed.
To return to the user mode, hold down the
“
MENU
”
key for 3 seconds or more.
[Note]
Setting all the channels to Skip displays an error,
.
At least, one channel should be set to measurement.
The skip channel conducts measurement, but does not provide display, print, or alarm judgment. If input is not
connected, short-circuit a measurement terminal.
It is not possible to return it based on the setting skipped once. Please set range again.
MENU
ESC
3sec
ENT
ENT
ENT
ENT
Display
Operation keys
Description