To access the internal components of the Moblvac

®

 III-cs, remove the two (2) Phillips head screws lo-

cated in the rear of the unit near the back assembly.  Tilt the cover forward. 

 
REVENTATIVE MAINTENANCE 

 
Preventative maintenance is recommended every six months.  It is up to the user’s discretion to clean 
the unit more often after frequent use or to lengthen the schedule if use is infrequent. 
 
A Preventative Maintenance Kit is available which contains all necessary items and instructions for 
performing the  preventative maintenance procedure. 
 

VACUUM REGULATOR REPLACEMENT 

 
1. 

Tilt the shroud assembly forward off of the Moblvac

®

 III-cs base (make note of the orientation 

of all tubing and the regulator body). 

2. 

Disconnect all three (3) tubing connections from the regulator body. 

3. 

Use a small flathead screwdriver to loosen the set screw on the vacuum regulator knob.   Un-
screw and remove the vacuum regulator knob from the control panel. 

4. 

Remove the outside locknut that secures the regulator to the control panel.  The regulator can 
now be removed. (Please note:  the depth of the inside locknut on the shaft of the regulator 
body. This determines the height of the regulator body in the control panel). 

5. 

To mount the new regulator, first remove the knob from the body by loosening the set screw on 
the vacuum regulator knob (Use a flathead screwdriver). 

6. 

Mount the new regulator body to the control panel making sure the fittings are in the same ori-
entation as the original, by re-installing the outside locknut. 

7. 

Insert the regulator knob fully onto regulator adjusting shaft. 

8. 

Tighten the set screw until regulator knob is secure to shaft and cannot be pulled off by hand. 

9. 

Reconnect all tubing.  

 

VACUUM REGULATOR CLEANING   

 
1. 

Remove the vacuum regulator assembly as described in the VACUUM REGULATOR RE-
PLACEMENT section above. 

Содержание Moblvac III-cs

Страница 1: ...CONSTANT SUCTION UNIT OPERATOR MAINTENANCE M A N U A L Ohio Medical Corporation 1111 Lakeside Drive Gurnee Illinois 60031 4099 1 800 662 5822 847 855 0800 Fax 847 855 6218 www ohiomedical com...

Страница 2: ...es Because of the Moblvac s constant suction capacities are driven by a fan cooled rotary carbon vane pump It comes equipped with a disposable exhaust filter an in line suction filter with 14 of tubin...

Страница 3: ...w located at the back of the base assembly with a 3 32 Allen wrench 5 Connect the tubing from the bottom of the backpole to the safety overflow trap assembly 6 Plug the power cord into an electrical o...

Страница 4: ...Moblvac III cs to the vacuum port of the collection canister and from the patient port of the collection canister to the aspirating tip that is being used 2 Open the vacuum port located at the top of...

Страница 5: ...port located at the top of the backpole Adjust the Ohio Medical needle valve or Chest Drainage Unit per the manufacturers instructions as necessary and proceed with the suction procedure REPROCESSING...

Страница 6: ...ng 2 Clean the overflow trap assembly by unscrewing the cap and disassembling the various compo nents Wash all parts with mild antiseptic dry reassemble and verify that the float is in the down positi...

Страница 7: ...nections from the regulator body 3 Use a small flathead screwdriver to loosen the set screw on the vacuum regulator knob Un screw and remove the vacuum regulator knob from the control panel 4 Remove t...

Страница 8: ...e regulator in the Moblvac III cs 8 Attach new vacuum tubing PUMP REPLACEMENT 1 Disconnect the three 3 pump wires from the PCB and two 2 wires from the capacitor 2 Disconnect the tubing from the fitti...

Страница 9: ...he three 3 pump head screws 7 Reinstall the pump assembly 3 PERFORMANCE TEST AND ADJUSTMENT It is recommended that you verify performance or the Moblvac III cs after An overflow Preventative maintenan...

Страница 10: ...the unit into an outlet 2 Faulty electrical connections 2 Make sure that all wires are secured tightly on the lugs and the lugs themselves are secure on the terminals 3 Pump has seized 3 Clean pump a...

Страница 11: ...Hg ELECTRICAL REQUIREMENTS AC 120V 60Hz 2A 220V 50Hz 2A Fuses Slow blow COLLECTION DEVICE Canister Disposable plastic with mechanical shutoff Capacity 1200cc standard Tubing 14 with bacterial filter P...

Страница 12: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 12 FIGURE 1...

Страница 13: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 13 Figure 2...

Страница 14: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 14 FIGURE 3...

Страница 15: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 15...

Страница 16: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 16 FIGURE 4...

Страница 17: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 17 FIOGURE 5...

Страница 18: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 18 FIGURE 6...

Страница 19: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 19...

Страница 20: ...ion 800 662 5822 Fax 847 855 6218 www aerosinstruments com 550107 Rev 5 08 2005 1111 Lakeside Drive Gurnee IL 60031 4099 Phone 800 662 5822 847 855 0800 Fax 847 855 6218 Email info ohiomedical com htt...

Отзывы: