Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218   

www.ohiomedical.com  —–  550107 (Rev.5) 08/2005 

2. Discard 

all 

tubing 

3. 

Open the regulator by removing the two screws on the cap, using a phillips screwdriver. 

4. 

Disassemble black cartridge from unit by unscrewing it from the brass adjustment screw and 
cap sub-assembly and remove two o-rings. 

5. 

Clean the body, adjusting screw and cap, fittings and O-rings in a mild soap solution or isopro-
pyl alcohol and dry all parts completely.  Do not wash black cartridge. 

6. 

Apply a light coating of a silicone based lubricant (i.e.: Dow Corning 111) to the O-rings and 
reassemble back onto the black cartridge. 

7. 

Reassemble the vacuum regulator and remount the regulator in the Moblvac

®

 III-cs. 

8. 

Attach new vacuum tubing.   

 

PUMP REPLACEMENT 

 
1. 

Disconnect the three (3) pump wires from the PCB and two (2) wires from the capacitor. 

2. 

Disconnect the tubing from the fittings on the pump. 

3. 

If necessary, loosen the mounting bracket for the capacitor and remove the capacitor. 

4. 

Using an 11/32” nut driver, remove the four (4) nuts from the base of the pump.  Remove 
pump. 

5. 

To mount replacement pump, insert the pump onto the four (4) vibration bumper posts and se-
cure the pump in place. 

6. 

Connect the pump wires to the PCB and the capacitor. 

7. 

Reconnect all tubing. 

8. 

Proceed to PERFORMANCE TEST & ADJUSTMENTS on page 9. 

 
PUMP CLEANING 

 

NOTE:

 The heart of the Moblvac

®

 III-cs is a rotary carbon vane pump.  It is NOT recommended that 

the pump be disassembled for routine cleaning.  However, if performance has been affected by the 
pump becoming contaminated with aspirant, or if an overflow problem has occurred, the following 

Содержание Moblvac III-cs

Страница 1: ...CONSTANT SUCTION UNIT OPERATOR MAINTENANCE M A N U A L Ohio Medical Corporation 1111 Lakeside Drive Gurnee Illinois 60031 4099 1 800 662 5822 847 855 0800 Fax 847 855 6218 www ohiomedical com...

Страница 2: ...es Because of the Moblvac s constant suction capacities are driven by a fan cooled rotary carbon vane pump It comes equipped with a disposable exhaust filter an in line suction filter with 14 of tubin...

Страница 3: ...w located at the back of the base assembly with a 3 32 Allen wrench 5 Connect the tubing from the bottom of the backpole to the safety overflow trap assembly 6 Plug the power cord into an electrical o...

Страница 4: ...Moblvac III cs to the vacuum port of the collection canister and from the patient port of the collection canister to the aspirating tip that is being used 2 Open the vacuum port located at the top of...

Страница 5: ...port located at the top of the backpole Adjust the Ohio Medical needle valve or Chest Drainage Unit per the manufacturers instructions as necessary and proceed with the suction procedure REPROCESSING...

Страница 6: ...ng 2 Clean the overflow trap assembly by unscrewing the cap and disassembling the various compo nents Wash all parts with mild antiseptic dry reassemble and verify that the float is in the down positi...

Страница 7: ...nections from the regulator body 3 Use a small flathead screwdriver to loosen the set screw on the vacuum regulator knob Un screw and remove the vacuum regulator knob from the control panel 4 Remove t...

Страница 8: ...e regulator in the Moblvac III cs 8 Attach new vacuum tubing PUMP REPLACEMENT 1 Disconnect the three 3 pump wires from the PCB and two 2 wires from the capacitor 2 Disconnect the tubing from the fitti...

Страница 9: ...he three 3 pump head screws 7 Reinstall the pump assembly 3 PERFORMANCE TEST AND ADJUSTMENT It is recommended that you verify performance or the Moblvac III cs after An overflow Preventative maintenan...

Страница 10: ...the unit into an outlet 2 Faulty electrical connections 2 Make sure that all wires are secured tightly on the lugs and the lugs themselves are secure on the terminals 3 Pump has seized 3 Clean pump a...

Страница 11: ...Hg ELECTRICAL REQUIREMENTS AC 120V 60Hz 2A 220V 50Hz 2A Fuses Slow blow COLLECTION DEVICE Canister Disposable plastic with mechanical shutoff Capacity 1200cc standard Tubing 14 with bacterial filter P...

Страница 12: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 12 FIGURE 1...

Страница 13: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 13 Figure 2...

Страница 14: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 14 FIGURE 3...

Страница 15: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 15...

Страница 16: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 16 FIGURE 4...

Страница 17: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 17 FIOGURE 5...

Страница 18: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 18 FIGURE 6...

Страница 19: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 19...

Страница 20: ...ion 800 662 5822 Fax 847 855 6218 www aerosinstruments com 550107 Rev 5 08 2005 1111 Lakeside Drive Gurnee IL 60031 4099 Phone 800 662 5822 847 855 0800 Fax 847 855 6218 Email info ohiomedical com htt...

Отзывы: