procedure should be performed. 
 
1. 

Remove the pump assembly from the Moblvac III-cs as described above in the PUMP  

 REPLACEMENT 

section. 

2. 

.Remove the three (3) pump head screws located on the pump head. 

 
3. 

Remove the cover plate, shim, wear plate and the four (4) vanes. 

4. 

Wash all exposed areas with an isopropyl alcohol solution. 

 
Before you reassemble the pump make certain that all components are completely dry.  Rust is likely 
to form if any moisture is present. 
 
5. 

Replace the vanes, wear plate, shim, and cover plate. 

 
6. 

Replace and equally tighten the three (3) pump head screws. 

 
7. 

Reinstall the pump assembly. 

 
 

3.  PERFORMANCE TEST AND ADJUSTMENT 

 
It is recommended that you verify performance or the Moblvac III-cs after: 
 
    - 

An 

overflow. 

    - 

Preventative 

maintenance. 

    - 

Any 

maintenance. 

 

To verify pump & regulator operation: 

 
1. 

Plug the power cord into an electrical outlet.  Turn the unit ON and listen to verify that the 
pump starts. 

 
2. 

Open the vacuum port located at the top of the backpole assembly and place your finger over 
the outlet. 

 
3. 

Adjust the vacuum level by turning the vacuum regulator knob.  Verify that the vacuum gauge 
reflects a change in vacuum level while turning the regulator knob.  Also verify that you feel 
vacuum at your finger tip. 

 
4. 

Turn the unit OFF. 

 

The following steps will verify the flow and the vacuum specifications: 

 
1. 

Connect a suction flow measuring device to the vacuum outlet of the Moblvac

®

 III-cs. 

 
2. 

Turn the unit 

ON

Содержание Moblvac III-cs

Страница 1: ...CONSTANT SUCTION UNIT OPERATOR MAINTENANCE M A N U A L Ohio Medical Corporation 1111 Lakeside Drive Gurnee Illinois 60031 4099 1 800 662 5822 847 855 0800 Fax 847 855 6218 www ohiomedical com...

Страница 2: ...es Because of the Moblvac s constant suction capacities are driven by a fan cooled rotary carbon vane pump It comes equipped with a disposable exhaust filter an in line suction filter with 14 of tubin...

Страница 3: ...w located at the back of the base assembly with a 3 32 Allen wrench 5 Connect the tubing from the bottom of the backpole to the safety overflow trap assembly 6 Plug the power cord into an electrical o...

Страница 4: ...Moblvac III cs to the vacuum port of the collection canister and from the patient port of the collection canister to the aspirating tip that is being used 2 Open the vacuum port located at the top of...

Страница 5: ...port located at the top of the backpole Adjust the Ohio Medical needle valve or Chest Drainage Unit per the manufacturers instructions as necessary and proceed with the suction procedure REPROCESSING...

Страница 6: ...ng 2 Clean the overflow trap assembly by unscrewing the cap and disassembling the various compo nents Wash all parts with mild antiseptic dry reassemble and verify that the float is in the down positi...

Страница 7: ...nections from the regulator body 3 Use a small flathead screwdriver to loosen the set screw on the vacuum regulator knob Un screw and remove the vacuum regulator knob from the control panel 4 Remove t...

Страница 8: ...e regulator in the Moblvac III cs 8 Attach new vacuum tubing PUMP REPLACEMENT 1 Disconnect the three 3 pump wires from the PCB and two 2 wires from the capacitor 2 Disconnect the tubing from the fitti...

Страница 9: ...he three 3 pump head screws 7 Reinstall the pump assembly 3 PERFORMANCE TEST AND ADJUSTMENT It is recommended that you verify performance or the Moblvac III cs after An overflow Preventative maintenan...

Страница 10: ...the unit into an outlet 2 Faulty electrical connections 2 Make sure that all wires are secured tightly on the lugs and the lugs themselves are secure on the terminals 3 Pump has seized 3 Clean pump a...

Страница 11: ...Hg ELECTRICAL REQUIREMENTS AC 120V 60Hz 2A 220V 50Hz 2A Fuses Slow blow COLLECTION DEVICE Canister Disposable plastic with mechanical shutoff Capacity 1200cc standard Tubing 14 with bacterial filter P...

Страница 12: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 12 FIGURE 1...

Страница 13: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 13 Figure 2...

Страница 14: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 14 FIGURE 3...

Страница 15: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 15...

Страница 16: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 16 FIGURE 4...

Страница 17: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 17 FIOGURE 5...

Страница 18: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 18 FIGURE 6...

Страница 19: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 19...

Страница 20: ...ion 800 662 5822 Fax 847 855 6218 www aerosinstruments com 550107 Rev 5 08 2005 1111 Lakeside Drive Gurnee IL 60031 4099 Phone 800 662 5822 847 855 0800 Fax 847 855 6218 Email info ohiomedical com htt...

Отзывы: