The Moblvac

®

 III-cs will use any collection device.  However, if you are not using an Ohio Medical 

1200cc disposable   collection canister, make sure that the collection device that is being used is 
equipped with a safety overflow mechanism to protect the pump from accidental overflow. 
 
Ensure that you have the appropriate collection device and suction tubing.  Depending on the proce-
dure, ensure that you have the appropriate Chest Drainage Unit, Nasal Gastric tube, aspirating tip, or 
wound drain for patient use.  The Moblvac

®

 III-cs is now ready to place into service. 

 

OPERATING INSTRUCTIONS 

 
Verify that a clean bacteriostatic exhaust filter, in-line suction filter, collection canister, and the neces-
sary tubing are installed on the Moblvac

®

 III-cs. 

 

ORAL, NASAL AND TRACHEAL ASPIRATION 

 
1. 

Place a clean collection canister in the upper bracket, located on the backpole assembly of your  
Moblvac

® 

III-cs. Ensure that tubing is connected from the “VACUUM PORT ASSEMBLY” of 

the Moblvac

®

 III-cs to the “vacuum port” of the collection canister and from the “patient port” 

of the collection  canister to the aspirating tip that is being used. 

 
2. 

Open the vacuum port located at the top of the backpole assembly and place your finger over 
the outlet. 

 
3. 

Turn the Moblvac

®

 III-cs 

ON

 
3. 

Adjust the vacuum to the desired vacuum level by using the “VACUUM REGULATOR” 
Knob. 

 
5. 

Open the vacuum port located at the top of the backole and proceed with the suction procedure.  

 

 

 

PLEURAL DRAINAGE 

 
Moblvac

®

 III-cs can be utilized with any “Disposable Chest Drainage Unit” available on the market.  

Some chest drainage units come with a built in flow adjustment valve and some do not.  The Ohio 
Medical Needle Valve Assembly is available for those units that do not offer this feature. 
 
Follow all manufacturers’ directions on the Chest Drainage Unit for its setup and use. 
 
Chest Drainage Units Without Built In Flow Adjustment Valves. 
 
Place the Chest Drainage Unit in the “optional” Ohio Medical Chest Drainage Unit Bracket.  Ensure 
that tubing is connected from the “VACUUM PORT ASSEMBLY” of the Moblvac III-cs to the Ohio 
Medical Needle Valve Assembly which is mounted in the collection canister bracket of the backpole.  
Connect tubing from the Needle Valve Assembly to the vacuum port tubing of the Chest Drainage 
Unit. 

 

Содержание Moblvac III-cs

Страница 1: ...CONSTANT SUCTION UNIT OPERATOR MAINTENANCE M A N U A L Ohio Medical Corporation 1111 Lakeside Drive Gurnee Illinois 60031 4099 1 800 662 5822 847 855 0800 Fax 847 855 6218 www ohiomedical com...

Страница 2: ...es Because of the Moblvac s constant suction capacities are driven by a fan cooled rotary carbon vane pump It comes equipped with a disposable exhaust filter an in line suction filter with 14 of tubin...

Страница 3: ...w located at the back of the base assembly with a 3 32 Allen wrench 5 Connect the tubing from the bottom of the backpole to the safety overflow trap assembly 6 Plug the power cord into an electrical o...

Страница 4: ...Moblvac III cs to the vacuum port of the collection canister and from the patient port of the collection canister to the aspirating tip that is being used 2 Open the vacuum port located at the top of...

Страница 5: ...port located at the top of the backpole Adjust the Ohio Medical needle valve or Chest Drainage Unit per the manufacturers instructions as necessary and proceed with the suction procedure REPROCESSING...

Страница 6: ...ng 2 Clean the overflow trap assembly by unscrewing the cap and disassembling the various compo nents Wash all parts with mild antiseptic dry reassemble and verify that the float is in the down positi...

Страница 7: ...nections from the regulator body 3 Use a small flathead screwdriver to loosen the set screw on the vacuum regulator knob Un screw and remove the vacuum regulator knob from the control panel 4 Remove t...

Страница 8: ...e regulator in the Moblvac III cs 8 Attach new vacuum tubing PUMP REPLACEMENT 1 Disconnect the three 3 pump wires from the PCB and two 2 wires from the capacitor 2 Disconnect the tubing from the fitti...

Страница 9: ...he three 3 pump head screws 7 Reinstall the pump assembly 3 PERFORMANCE TEST AND ADJUSTMENT It is recommended that you verify performance or the Moblvac III cs after An overflow Preventative maintenan...

Страница 10: ...the unit into an outlet 2 Faulty electrical connections 2 Make sure that all wires are secured tightly on the lugs and the lugs themselves are secure on the terminals 3 Pump has seized 3 Clean pump a...

Страница 11: ...Hg ELECTRICAL REQUIREMENTS AC 120V 60Hz 2A 220V 50Hz 2A Fuses Slow blow COLLECTION DEVICE Canister Disposable plastic with mechanical shutoff Capacity 1200cc standard Tubing 14 with bacterial filter P...

Страница 12: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 12 FIGURE 1...

Страница 13: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 13 Figure 2...

Страница 14: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 14 FIGURE 3...

Страница 15: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 15...

Страница 16: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 16 FIGURE 4...

Страница 17: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 17 FIOGURE 5...

Страница 18: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 18 FIGURE 6...

Страница 19: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 19...

Страница 20: ...ion 800 662 5822 Fax 847 855 6218 www aerosinstruments com 550107 Rev 5 08 2005 1111 Lakeside Drive Gurnee IL 60031 4099 Phone 800 662 5822 847 855 0800 Fax 847 855 6218 Email info ohiomedical com htt...

Отзывы: