CHAPTER 3 MAINTENANCE PROCEDURES
Ohaus Corporation www.ohaus.com
3-12
MB23 and MB25 Service Manual
3.8.7 Replacing the Thermal Overload Device
1. Remove the Chamber Top and
Reflector (Sections 3.9.1 and 3.9.2).
2. Remove the cables from the Thermal
Overload Device by pressing on the
tiny spring with the end of a small flat
head screwdriver. (See Figure 3-21.)
3. If the Thermal Overload Device has
been disassembled, insert its bolt
through the hole in the base of its
blade. (See Figure 3-27.)
4. Fit the washer and then the nut, and
use a wrench to tighten the nut.
Figure 3-27. MB25 Thermal Overload
Device.
5.
MB23:
Mount the ground cable ring.
(See Figure 3-28.) Then use a wrench
to tighten the nut.
Figure 3-28. MB23 Thermal Overload
Device.
6. Insert Thermal Overload Device into
the grooves provided on the Housing.
7. Connect the cables to the Thermal
Overload as shown in Figure 3-29.
7. Reassemble the Heating Unit.
8. Perform a Temperature Calibration.
(See Appendix A.)
9. Conduct Performance Tests.
(See Chapter 4.)
Figure 3-29. MB25 Thermal Overload
installed.
Содержание MB25
Страница 1: ...SERVICE MANUAL Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA 973 377 9000...
Страница 2: ......
Страница 4: ......
Страница 8: ...TABLE OF CONTENTS MB23 and MB25 Service Manual iv Ohaus Corporation www ohaus com...
Страница 43: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 4 MB23 and MB25 Service Manual...
Страница 47: ...APPENDIX A STANDARD CALIBRATION Ohaus Corporation www ohaus com A 4 MB23 and MB25 Service Manual...