CHAPTER 2 TROUBLESHOOTING
Ohaus Corporation www.ohaus.com
2-2
MB23 and MB25 Service Manual
2.2 Error Codes
When a problem occurs, the display may indicate an error code. Table 2-1 describes the
various error codes and specifies the probable cause and remedy.
TABLE 2-1. ERROR CODES
ERROR
PROBABLE CAUSE
REMEDY
Err 7.0
Time out
, the weighing value is not stable
Defective Main PCB.
Turn unit off and then back on.
Replace Main PCB. Section 3.6.
Err 8.1
Pan Support has load during power on.
Dirt or debris in unit.
Defective Load Cell.
Defective Main PCB.
Remove weight from Pan Support.
Clean the unit thoroughly.
Replace Load Cell. Section 3.7.
Replace Main PCB. Section 3.6.
Err 8.2
Pan Support removed during power on, or dirty
Pan Support.
Load Cell defective.
Main PCB defective.
Install Pan Support.
Be sure it’s clean.
Replace Load Cell. Section 3.7.
Replace Main PCB. Section 3.6.
Err 8.3
Weight on Pan Support exceeds capacity.
Dirt or debris in unit.
Remove weight from Pan Support.
Perform troubleshooting steps. Section 2.1.
Clean the unit thoroughly.
Err 8.4
Pan Support was removed during weighing.
Dirt or debris in unit.
Re-install Pan Support.
Perform troubleshooting steps. Section 2.1.
Clean the unit thoroughly.
Err 8.5
Analyzer temperature is higher than 50°C
Cool down the analyzer
Err 8.6
1. Analyzer temperature is higher than target
setting. (20°C higher than target setting)
2. Bad temperature calibration
3. Temperature Sensor defective
Cool down the analyzer
Perform temperature calibration
Replace temperature sensor
Err 9.5
Factory calibration data corrupted.
Use Software Service Tool to restore EEPROM.
Appendix B.
Err 10.3
1. Bad temperature calibration
2. Temperature Sensor defective
Check/Replace cables. Section 3.3.2.
Check/Replace Sensor. Section 3.8.9.
Replace Main PCB. Section 3.6.
Perform temperature calibration
Err 10.4
Temperature Sensor out of range
– low
Check/Replace cables. Section 3.3.2.
Check/Replace Sensor. Section 3.8.9.
Replace Main PCB. Section 3.6.
Err 10.5
Temperature Sensor defective
Check/Replace cables. Section 3.3.2.
Check/Replace Sensor. Section 3.8.9.
Replace Main PCB. Section 3.6.
Err 10.6
Temperature Sensor defective
Check/Replace cables. Section 3.3.2.
Check/Replace Sensor. Section 3.8.9.
Replace Main PCB. Section 3.6.
Err 10.7
Temperature out of control (Current temperature is
20°C higher than target setting)
Use a Power Supply Regulator
Replace temperature Sensor Section 3.8.9.
Replace main PCB. Section 3.6.
Содержание MB25
Страница 1: ...SERVICE MANUAL Ohaus Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA 973 377 9000...
Страница 2: ......
Страница 4: ......
Страница 8: ...TABLE OF CONTENTS MB23 and MB25 Service Manual iv Ohaus Corporation www ohaus com...
Страница 43: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 4 MB23 and MB25 Service Manual...
Страница 47: ...APPENDIX A STANDARD CALIBRATION Ohaus Corporation www ohaus com A 4 MB23 and MB25 Service Manual...