EN-12
EXPLORER BALANCES
2.8.2
Terminal Mounting
If desired, the Terminal may be mounted to a wall or table using fasteners (not supplied) that are appropriate for the
type of mounting surface The Terminal bracket will accommodate 4mm (#8) screws. Locate the mounting holes as
shown in Figure 2-1.
Figure 2-1. Terminal Mounting Dimensions.
2.9
Initial Calibration
When the Balance is first installed, and when it is moved to another location, it must be calibrated to ensure
accurate weighing results. The Explorer Balances have built in AutoCal which can calibrate the balance
automatically and does not require calibration masses. If preferred, the balance can be manually calibrated with
external masses. Have the appropriate calibration masses available before beginning calibration. Refer to the
Calibration Section for masses and calibration procedure.
56.5 mm
2.2 in
112 mm / 4.4 in
48.5 mm
1.9 in