EN-96
EXPLORER BALANCES
5.5
Balance Setup
Enter this sub-menu to customize Balance display functionality.
5.5.1
Balance Set-up sub-menu
Stability
Indicator Range
Filter Level
Auto Zero
Tracking
Auto Tare
Gross Indicator
Legal For Trade
Graduations
Date & Time
Factory default settings are shown below in bold.
5.5.2
Stable Range
Set the amount the reading can vary while the stability symbol remains on.
0.5 D
= 0.5 graduations
1 D
= 1 graduation
2 D
= 2 graduations
3 D
= 3 graduations
5 D
= 5 graduations
Note:
When LEGAL FOR TRADE (below) is set to ON, the setting is forced to 1 D.
The setting is locked when the Security Switch is set to the ON position.
5.5.3
Filter level
Set the amount of signal filtering.
LOW
= faster stabilization time with less stability.
MEDIUM
= normal stabilization time with normal stability.
HIGH
= slower stabilization time with more stability.
5.5.4
Auto Zero Tracking
Set the automatic zero tracking functionality.
OFF
= disabled.
0.5 D
= display maintains zero up to a drift of 0.5 graduation
per second
1 D
= display maintains zero up to a drift of 1 graduation per
second.
3 D
= display maintains zero up to a drift of 3 graduations per
second.
Note:
When Legal for Trade is set to ON, the AZT setting is forced to 0.5 D.
The OFF setting is still available. The setting is locked at the current setting when the
security switch is set to the ON position