Defender 3000 Indicators Service Manual
EN-11
1.7 Internal Connections
Some internal connections require to open the housing first. Therefore, please check the
Open the
Housing
section below before making connections.
CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO
THE INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. THE
HOUSING SHOULD ONLY BE OPENED BY AUTHORIZED AND QUALIFIED PERSONNEL,
SUCH AS AN ELECTRICAL TECHNICIAN.
1.7.1 Load cell connection EasyConnect
TM
Indicator
To connect an OHAUS EasyConnect
TM
scale base to an i-DT33P or an i-DT33XW EasyConnect
TM
indicator,
plug the base's connector into the external load cell connector located at the back (i-DT33P) or bottom (i-
DT33XW) of the indicator. Then rotate the base connector’s locking ring in the direction of the arrows
demonstrated below.
Note
: For connecting bases which do not support EasyConnect
TM
to an EasyConnect
TM
indicator, contact
an authorized OHAUS dealer to obtain a load cell cable gland kit (P/N 30379716) as an accessory.
1.7.2 Load cell connection Non-EasyConnect
TM
Indicator
To connect a non-EasyConnect
TM
scale base to an i-DT33P or an i-DT33XW non-EasyConnect
TM
indicator,
follow the instruction in
Open the Housing
section to open the housing first before proceeding.
Note
: For connecting EasyConnect
TM
bases to a non-EasyConnect
TM
indicator, contact an authorized
OHAUS dealer to obtain a load cell cable gland kit (P/N 30427858) as an accessory.
i-DT33P
i-DT33XW
Содержание Defender 3000 Series
Страница 1: ...Defender 3000 Indicators Service Manual i DT33P i DT33XW...
Страница 2: ......
Страница 9: ...Defender 3000 Indicators Service Manual EN 7 1 4 3 Main PCB...
Страница 10: ...EN 8 Defender 3000 Indicators Service Manual 1 5 Control Functions i DT33P Control Panel i DT33XW Control Panel...
Страница 11: ...Defender 3000 Indicators Service Manual EN 9...
Страница 50: ......