EN-44
Defender 3000 Indicators Service Manual
APPENDIX A. CONFIGURING THE MAIN PCBA
Note:
To perform the configuration, connect the indicator with computer via RS232. Please refer to section
1.7.3 for i-DT33P, section 1.7.4 for i-DT33XW for the RS232 connection.
Once a new Main PCBA is installed, you need to perform PCB configure according to their region.
1. Down load and install OHAUS Service and Repair Tools (Version 2.4.3.3 and later).
2. Open the software and select 'Repair Tool' and click 'Next'
3. Select 'DEFENDER i-DT33' under Product Selection and click 'Next'.
4. Select 'Replace Main PCB', click 'Next'
5. Set the same protocol setting between indicator and computer and click 'Next'.
6. Input the 'Serial Number' and 'Model Number' , Select the 'Shipping Country' from the location list
and click 'Next'.
7. You will see the below window screen when the configuration is completed successfully.
8. Turn OFF and ON the indicator and check if the configuration is correct.
9. If the process fail kindly check the below.