Defender 3000 Indicators Service Manual 

 

EN-21 

 

2.3.3 Span Calibration 

[

SpaN

]

 

Span calibration uses one point. The span calibration point is established with a calibration mass placed on 
the scale. 

Note:

 Span calibration should be performed after zero calibration. 

Calibration procedures: 

1. 

Long press the 

Menu

 button until you see 

mM.E.N.U

.  

2. 

Release the button and wait for the display to show 

C.A.L

. Press the 

Yes

 button.  

3. 

Short press the 

No

 button to navigate until you see 

SPAN

.

 

Press the 

Yes

 button.

 

4. 

The calibration point and the unit are flashing on the display based on the scale capacity and unit set in 

the menu. (e.g. 

030.000

 kg) If you do not need to change the calibration point, skip to step 6.  

5. 

To change the calibration point: 
a) 

Short press the 

No

 button several times until the desired number appears. 

Note

: press the 

Back

 button can decrease the digit.  

b) 

Short press the 

Yes

 button to accept the number and move on to the next digit.  

c) 

Repeat the process until all the digits are correct.  

d) 

Press the 

Yes

 button to accept calibration point. It is flashing on the display. 

6. 

Place a calibration mass of the specified weight on the pan and press the 

Yes

 button.  

7. 

The display shows 

--C--

, and then

 -DONE- 

when the calibration is finished.  

8. 

Then the display shows 

Lin

. Press the 

Exit

 button to exit.  

Note:  

  If calibration is failed, 

CAL E 

is displayed for 3 seconds and the previous calibration data is 

restored.  

  If after waiting for 40 seconds the calibration is still not successful, 

CAL E

 is displayed for 3 seconds 

and the previous calibration data is restored.  

 

2.3.4 Linearity Calibration 

[

LIN

]

 

Linearity calibration uses 3 calibration points. The full calibration point is established with a weight on the 
scale. The mid calibration point is established with a weight equal to half of the full calibration weight on the 
scale. The zero calibration point is established with no weight on the scale. The full calibration and mid 
calibration points can be altered by the user during the calibration procedure. 

Calibration procedures: 

1. 

Long press the 

Menu

 button until you see 

mM.E.N.U

.  

2. 

Release the button and wait for the display to show 

C.A.L

. Press the 

Yes

 button.  

3. 

Short press the 

No

 button several times to navigate until you see 

LIN

.

 

Press the 

Yes

 button.

 

4. 

0

 kg and the calibration unit are flashing on the display. With no weight on the pan, press the 

Yes

 

button to establish the zero point. The display shows 

--C--

5. 

Then the first calibration point and the unit are flashing on the display based on the scale capacity and 

unit you set in the setup menu. (For example, 

015.000

 kg). If you do not need to change the calibration 

point, skip to step 7. 

6. 

To change the calibration point: 
a) 

Short press the 

No

 button several times until the desired number appears. 

Note

: press the 

Back

 button can decrease the digit.  

b) 

Short press the 

Yes

 button to accept the number and move on to the next digit.  

c) 

Repeat the process until all the digits are correct.  

d) 

Press the 

Yes

 button to accept calibration point. It is flashing on the display. 

7. 

Place a calibration mass of the specified weight on the pan and press the 

Yes

 button. The display 

shows  

--C--

Содержание Defender 3000 Series

Страница 1: ...Defender 3000 Indicators Service Manual i DT33P i DT33XW...

Страница 2: ......

Страница 3: ...ternal Connections 10 1 6 1 Battery Power Supply 10 1 6 2 AC Power to i DT61XWE 10 1 7 Internal Connections 11 1 7 1 Load cell connection EasyConnectTM Indicator 11 1 7 2 Load cell connection Non Easy...

Страница 4: ...Information 26 4 TECHNICAL DATA 27 4 1 Specifications 27 4 2 Table of Geo Values 29 5 REPLACING MAJOR COMPONENTS INDICATOR 30 5 1 Open the Housing 30 5 2 Printed Circuit Board PCB Replacement 32 5 3 P...

Страница 5: ...operty or in loss of data or minor or medium injuries if not avoided ATTENTION For important information about the product May lead to equipment damage if not avoided NOTE For useful information about...

Страница 6: ...tute a hazard in use If the instrument is not used according to these operating instructions the intended protection provided by the instrument may be impaired 1 2 Service Facilities To service an ind...

Страница 7: ...testing the RS232 USB communications 4 RS232 Interface cable for i DT33XW 5 Data Printer for use with RS232 USB communications 1 3 2 Standard Tools and Test Equipment List 1 Standard Electronics Tool...

Страница 8: ...DT33XW Note There are two types of load cell connector one supports EasyConnectTM and one does not Therefore the load cell connector varies for different models For some i DT33XW models the position o...

Страница 9: ...Defender 3000 Indicators Service Manual EN 7 1 4 3 Main PCB...

Страница 10: ...EN 8 Defender 3000 Indicators Service Manual 1 5 Control Functions i DT33P Control Panel i DT33XW Control Panel...

Страница 11: ...Defender 3000 Indicators Service Manual EN 9...

Страница 12: ...when the battery is fully charged While operating on battery the battery symbol indicates the battery status Battery 5 25 remaining Battery 25 50 remaining Battery 50 75 remaining Battery 75 100 remai...

Страница 13: ...the external load cell connector located at the back i DT33P or bottom i DT33XW of the indicator Then rotate the base connector s locking ring in the direction of the arrows demonstrated below Note Fo...

Страница 14: ...sing release the load cell connector circled on the rear of the indicator and then pass the load cell cable through it 2 Connect all the cables on the load cell cable to the load cell terminal block o...

Страница 15: ...l EN 13 i DT33XW 1 After opening the housing unplug the battery connectors circled and take out the battery 2 Unscrew the two screws on the sealing cover circled take out the cover and then take out t...

Страница 16: ...tor circled on the bottom of the indicator and then pass the load cell cable through it 4 Connect all the cables on the load cell cable to the load cell terminal block on the mainboard circled Please...

Страница 17: ...tion EXE and Sense SEN terminals The i DT33P and i DT33XW indicators are designed to support both 2mV V and 3mV V load cells from the same circuitry A load cell output rating selection jumper is not r...

Страница 18: ...using section for how to open the rear housing 2 Take out the battery Please refer to the i DT33XW sub section of Non EasyConnectTM Indicator section for how to take out the battery 3 Release the stra...

Страница 19: ...table using fasteners not supplied that are appropriate for the type of mounting surface The bracket will accommodate up to 6 mm 1 4 diameter screws Locate the mounting holes as shown below RS232 ter...

Страница 20: ...oz t End reset BAUd PArItY StOP H SHAKE ALt P ALt t ALt Z End reset ASSIGN Stable mMODE tImME lINkto C SUmM tEmMp End reset BAUd PArItY StOP H SHAKE ALt P ALt t ALt Z ENd reset IP Addr u E r ENd reset...

Страница 21: ...For menu items which need to set numbers such as the scale capacity the current number is displayed with all digits flashing To revise 1 Press the No button to begin editing 2 The first digit is displ...

Страница 22: ...lb 120 lb 150 kg 150 kg 300 lb 300 lb Note The calibration unit is consistent with the capacity unit 2 3 2 Zero Calibration ZErO Zero calibration uses one calibration point The zero calibration point...

Страница 23: ...not successful CAL E is displayed for 3 seconds and the previous calibration data is restored 2 3 4 Linearity Calibration LIN Linearity calibration uses 3 calibration points The full calibration point...

Страница 24: ...the Back button can decrease the digit 6 Then the display shows C test Press the Exit button to exit 2 3 6 Calibration Test C test Calibration test procedures 1 Long press the Menu button until you se...

Страница 25: ...stant without fluctuations To view the Ramp value A number will appear and should be constant Place masses on the scale from minimum to maximum capacity The reading will increase but should not fluctu...

Страница 26: ...be operated Allow one hour for each 5 F 2 7 C temperature change before using the scale After temperature stabilization allow an additional 20 minutes after turning the scale on for the scale indicat...

Страница 27: ...been locked 1 Enable the Menu button in the L O C k Menu 2 The LFT security switch may need to be switched off Error 8 1 Weight reading exceeds Power On Zero limit 1 Remove weight from the pan 2 Reca...

Страница 28: ...th L and N pin should be higher than 20M or infinity Note The grounding pin of the model i DT33P to the metal surface of EasyconnectorTM should be infinity since it s mounted on the plastic housing bu...

Страница 29: ...EC OIML 1 6 000 NTEP Measurement Canada Class III Weighing units Kilogram Gram Pound Ounce Pound Ounce Tonne Metric Tonne Modes Weighing Counting Check Totalization Display 45 mm 1 8 in digit height...

Страница 30: ...unting Check Totalization Display 45 mm 1 8 in digit height LCD display with 4 color backlight Check indicator 3 color red green yellow bar Keyboard 4 button membrane keyboard Auto zero tracking Off 0...

Страница 31: ...15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 13 12 44 32 45 38 18 17 17 16 16 15 15 14 14 13 13 45 38 46 45 18 18 17 17 16 16 15 15 14 1...

Страница 32: ...xterior surfaces should be cleaned as soon as possible with warm water on a damp cloth Allow at least 10 minutes for the Indicator to stabilize after moving it from an area of a different temperature...

Страница 33: ...1 2 i DT33XW 1 Disconnect the Indicator from the AC power source 2 Remove the 6 bolts from the Bottom Plate 3 Disconnect the cable of battery and take out the battery carefully 4 Remove the 2 screws...

Страница 34: ...e the PCB for any of the following reasons Display is defective characters missing or partial display Indicator fails to calibrate properly Display is erratic or unstable Certain functions are not ope...

Страница 35: ...ON When handling the PCB grasp it by the edges only Do not touch the foil side Static discharge may damage some components 4 Reverse the process for the main PCB assembly 5 After replacing with a new...

Страница 36: ...ion label and loadcell from the main PCB 3 Remove the 3 screws and 1 copper hexagon nut securely remove the main PCB CAUTION When handling the PCB grasp it by the edges only Do not touch the foil side...

Страница 37: ...m the Power PCB to the Main PCBA 3 Remove the cable connector from the Power PCB to the AC Inlet connector 4 Remove the 4 screws and take out the Power PCB carefully 5 Reverse the process for the Powe...

Страница 38: ...ver of Power PCB 5 Remove the last screw and take out the Power PCB carefully 6 Reverse the process for the Power PCB assembly 5 4 Display Replacement 5 4 1 i DT33P 1 To open the housing please refer...

Страница 39: ...LCD carefully 5 Reverse the process for the New LCD display installation 5 4 2 i DT33XW 1 To open the housing please refer to section 5 1 2 2 Remove the main PCB section 5 2 2 and Power PCB section 5...

Страница 40: ...To open the housing refer to section 5 1 1 for i DT33P section 5 1 2 for i DT33XW 2 Disconnect the white cables connector from the main PCB and remove the quick connect nuts that hold the cables in pl...

Страница 41: ...trong adhesive 3 Clean the top cover and remove all traces of adhesive 4 Remove the protective backing from the back of the new function label Carefully position it on the top housing starting at the...

Страница 42: ...This section of the manual contains exploded views for the Defender 3000 Bases The exploded view drawings are designed to identify the parts which can be serviced in the field 6 1 Defender 3000 Series...

Страница 43: ...t Main PCBA i DT33 5 PCB Main AD7191 i DT33P PCB Main ADS1230 i DT33P 6 Switch Power Supply 7 Bottom Housing Kit i DT33P 8 Battery cover Kit i DT33P 9 Knob Kit i DT33 10 Bracket Kit U Bracket Assembly...

Страница 44: ...EN 42 Defender 3000 Indicators Service Manual 6 2 Defender 3000 Series Indicator i DT33XW Figure 6 2 Defender 3000 series Indicator i DT33XW...

Страница 45: ...Main AD7191 i DT33XW PCB Main ADS1230 i DT33XW 6 Switch Power Supply 7 Bottom Housing Kit i DT33XW 8 Hardware kits i DT33XW 9 Connector Kit i DT33XW 10 Rubber Kit i DT33XW 11 Loadcell Connector Short...

Страница 46: ...Repair Tools Version 2 4 3 3 and later 2 Open the software and select Repair Tool and click Next 3 Select DEFENDER i DT33 under Product Selection and click Next 4 Select Replace Main PCB click Next 5...

Страница 47: ...Defender 3000 Indicators Service Manual EN 45 The communication cable pin configuration The Indicator Protocol setting...

Страница 48: ...T33 under Product Selection and click Next 4 Select Download Software click Next 5 Set the same protocol setting between indicator and computer and click Next 6 Select the Mot file and click Next 7 Fo...

Страница 49: ...Indicators Service Manual EN 47 8 You will see the below window screen when it s in the process of the software download 9 You will see the below window screen when the software download is completed...

Страница 50: ......

Страница 51: ...S Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 1 973 377 9000 Fax 1 973 944 7177 With offices worldwide www ohaus com 30718420 P N 30718420A 2021 OHAUS Corporation all rights reser...

Отзывы: