Defender 3000 Indicators Service Manual
EN-21
2.3.3 Span Calibration
[
SpaN
]
Span calibration uses one point. The span calibration point is established with a calibration mass placed on
the scale.
Note:
Span calibration should be performed after zero calibration.
Calibration procedures:
1.
Long press the
Menu
button until you see
mM.E.N.U
.
2.
Release the button and wait for the display to show
C.A.L
. Press the
Yes
button.
3.
Short press the
No
button to navigate until you see
SPAN
.
Press the
Yes
button.
4.
The calibration point and the unit are flashing on the display based on the scale capacity and unit set in
the menu. (e.g.
030.000
kg) If you do not need to change the calibration point, skip to step 6.
5.
To change the calibration point:
a)
Short press the
No
button several times until the desired number appears.
Note
: press the
Back
button can decrease the digit.
b)
Short press the
Yes
button to accept the number and move on to the next digit.
c)
Repeat the process until all the digits are correct.
d)
Press the
Yes
button to accept calibration point. It is flashing on the display.
6.
Place a calibration mass of the specified weight on the pan and press the
Yes
button.
7.
The display shows
--C--
, and then
-DONE-
when the calibration is finished.
8.
Then the display shows
Lin
. Press the
Exit
button to exit.
Note:
If calibration is failed,
CAL E
is displayed for 3 seconds and the previous calibration data is
restored.
If after waiting for 40 seconds the calibration is still not successful,
CAL E
is displayed for 3 seconds
and the previous calibration data is restored.
2.3.4 Linearity Calibration
[
LIN
]
Linearity calibration uses 3 calibration points. The full calibration point is established with a weight on the
scale. The mid calibration point is established with a weight equal to half of the full calibration weight on the
scale. The zero calibration point is established with no weight on the scale. The full calibration and mid
calibration points can be altered by the user during the calibration procedure.
Calibration procedures:
1.
Long press the
Menu
button until you see
mM.E.N.U
.
2.
Release the button and wait for the display to show
C.A.L
. Press the
Yes
button.
3.
Short press the
No
button several times to navigate until you see
LIN
.
Press the
Yes
button.
4.
0
kg and the calibration unit are flashing on the display. With no weight on the pan, press the
Yes
button to establish the zero point. The display shows
--C--
.
5.
Then the first calibration point and the unit are flashing on the display based on the scale capacity and
unit you set in the setup menu. (For example,
015.000
kg). If you do not need to change the calibration
point, skip to step 7.
6.
To change the calibration point:
a)
Short press the
No
button several times until the desired number appears.
Note
: press the
Back
button can decrease the digit.
b)
Short press the
Yes
button to accept the number and move on to the next digit.
c)
Repeat the process until all the digits are correct.
d)
Press the
Yes
button to accept calibration point. It is flashing on the display.
7.
Place a calibration mass of the specified weight on the pan and press the
Yes
button. The display
shows
--C--
.
Содержание Defender 3000 Series
Страница 1: ...Defender 3000 Indicators Service Manual i DT33P i DT33XW...
Страница 2: ......
Страница 9: ...Defender 3000 Indicators Service Manual EN 7 1 4 3 Main PCB...
Страница 10: ...EN 8 Defender 3000 Indicators Service Manual 1 5 Control Functions i DT33P Control Panel i DT33XW Control Panel...
Страница 11: ...Defender 3000 Indicators Service Manual EN 9...
Страница 50: ......