EN-46
Defender 3000 Indicators Service Manual
APPENDIX B. SOFTWARE UPGRADE VIA SERVICE TOOL
Note:
To perform the software upgrade, connect the indicator with computer via RS232. Please refer to
section 1.7.3 for i-DT33P, section 1.7.4 for i-DT33XW for the RS232 connection.
1. Download and install OHAUS Service and Repair Tools (Version 2.4.3.3 and later).
2. Open the software and select 'Repair Tool' and click 'Next'
3. Select 'DEFENDER i-DT33' under Product Selection and click 'Next'.
4. Select 'Download Software', click 'Next'
5. Set the same protocol setting between indicator and computer and click 'Next'.
6. Select the Mot file and click 'Next'.
7. Follow by the instruction in the picture below, cycle the power.
Note:
Process to cycle the power as below:
1) Unplug the power supply for about 5 second
2) Press "On/Zero" to switch off the unit completely (this step for the models with battery)
3) Plug the power supply
4) Press "On/Zero" to switch on the unit (this step for the models
with
battery and the
models with
the main PCB 30707556 and both without the rechargeable PCB and
battery
)