Setup
Potentiometer Indicator
The HTHP Consistometer includes a built-in Eurotherm 2408i 15V
Potentiometer Indicator. As the slurry cup rotates within the unit, the poten-
tiometer creates a small voltage charge, which is displayed on the indicator.
This charge increases as the strength of the slurry increases.
The potentiometer indicator features a high-voltage alarm. When the
charge reaches the specified level, an alarm condition is triggered and the
heater, pump, and motor are automatically shut off.
To change the alarm setting, perform the following steps:
1. Push the “Page” key twice. The display will read “AL”.
2. Push the “Enter” key. The display will read “IFSH”.
3. Push the up arrow key once to read the current setting.
4. Push the up or down arrow keys to change the setting.
5. After choosing the appropriate setting, press the “Page” key three times
to return to the original display.
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
17