Setup
Chart Recorder
The OFITE HTHP Consistometer includes a Eurotherm Chessell 6100V
chart recorder for displaying and recording test data. It features a remov-
able drive for easily transferring test data to a PC for processing. The unit
is setup to automatically record data onto the disk during the test.
However, for this feature to work, you must have the disk inserted into the
drive when the unit is powered on. Otherwise, the data must be manually
archived at the end of the test.
It is strongly recommended that you carefully study the Eurotherm
Chessell 6100V instruction manual before using this equipment.
To manually archive test data:
1. Press the “Menu” button.
2. Press “Operator” from the menu.
3. At the top of the screen, press “Security”
and then choose “Log In”.
4. You will then be prompted for a UserID.
Touch the UserID field and choose
“Engineer” from the list.
5. Now touch the “Password” field. At the
bottom of the screen, choose “Numeric”
as the input method and then type “100”
and press “OK”.
6. At the top of the screen, choose “Archive”
and then choose “Local”.