9. Turn on the pump. Adjust the pressure to the desired level by turning
the regulator clockwise. When you reach the test pressure, turn off the
pump.
If the pressure rises too high, open (counter clockwise) the “Pressure
Release” valve very slowly. Close the valve immediately to prevent all
of the pressure from leaking.
10. Turn the heat on and push the “Run” button on the temperature con-
troller.
11. Turn the “Timer” switch on and push the “R” reset button on the timer
display.
The timer displays elapsed time to the nearest tenth of a minute. If an
alarm condition occurs, the timer will stop. This is a troubleshooting
measure that shows you the point during the test at which the alarm
occurred.
The “Timer” switch provides power to the timer and the “R” reset switch
resets the timer to zero.
12. If the “Alarm” switch is not already on, turn it on now. If the “Alarm”
switch is left off, the unit can still enter into an alarm condition, but there
will be no visual or audio signal to notify the operator.