OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
11
If for some reason the temperature controller fails, the thermostat will cut
power to the heater at 450°F (232.2°C), which is the maximum allowable
temperature.
Be careful in handling hot aging cells. Do not open cells while hot
or under pressure. Properly bleed off pressure and be sure the
valve stem is always pointed away from people or equipment. Al-
ways wear personal safety equipment.
6. Leave the “Heat” switch on the control panel ON until the oven and
sample have cooled down to handling temperature. The temperature
controller will indicate the oven temperature as long as the “Heat” switch
is on. For faster cooling, open the door.
7. After the test is complete, turn the “Heat” and “Motor” switches off.
Fuse for Heater
10 Amp –
115/230 Volt
Fuse for
Temp. Controller
1/2 Amp 115/230 Volt
Fuse for Motor
5 Amp – 115 Volt
2 Amp – 230 Volt
Rollers Indicator
Lamp - White
Motor Switch
Heat Indicator
Lamp - Red
Heat Switch
Control Panel