Computer interface
Note:
The commands SP2…SP6 concern the switching functions 2…6, re-
spectively. The commands are analogous to the SP1 command.
6.3.28
SPS
Setpoint status. Switching function status. See Chapter 5.1 Switching function
parameters (PArA SP),
S: SPS<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c,d,e,f<CR><LF>
b.bbbbE±bb
Lower threshold value in the current unit of mea-
surement
c.ccccE±cc
Upper threshold value in the current unit of mea-
surement
Parameters
Significance
a
Status of switching function 1
0 = Off
1 = On
b
Status of switching function 2
(see above)
c
Status of switching function 3
(see above)
d
Status of switching function 4
(see above)
e
Status of switching function 5
(see above)
f
Status of switching function 6
(see above)
Parameters
Significance