Parameters
5.2.6
Transmitter activation (
S-on
)
This parameter determines how the transmitter is switched on. The transmitter
activation can be set to one of the following values:
HAnd
Manually. The transmitter can be switched on by pressing the UP button. See
Chapter 4.4.3.3 Switching high-vacuum circuit on,
30.
Hot
Hot start. The transmitter automatically switches on when the unit is switched
on. After a power failure the measurement will be resumed automatically.
CH 1
By channel 1. The subsequent parameter
t-on
is used to specify the switch-on
threshold. The transmitter is switched on when the pressure on channel 1 falls
below the switch-on threshold.
CH 2
By channel 2. The subsequent parameter
t-on
is used to specify the switch-on
threshold. The transmitter is switched on when the pressure on channel 2 falls
below the switch-on threshold.
CH 3
By channel 3. This setting is only available if the unit is equipped with three
channels. The subsequent parameter
t-on
is used to specify the switch-on
threshold. The transmitter is switched on when the pressure on channel 3 falls
below the switch-on threshold.
5.2.7
Switch-on threshold (
t-on
)
This parameter is only available if the transmitter activation parameter is set to
CH 1
,
CH 2
or
CH 3
. See Chapter 5.2.6 Transmitter activation (S-on),
The
t-on
parameter is used to specify a switch-on threshold. The transmitter
is switched on when the pressure on the respective channel falls below the
switch-on threshold.
5.2.8
Transmitter deactivation (
S-oFF
)
This parameter determines how the transmitter is switched off. The transmitter
deactivation can be set to one of the following values:
HAnd
Manually. The transmitter can be switched off by pressing the DOWN button.
See Chapter 4.4.3.5 Switching high-vacuum circuit off,
31.
SELF
Self control. The subsequent parameter
t-off
is used to specify the switch-off
threshold. The transmitter is switched off when the pressure at the sensor
exceeds the switch-off threshold.
CH 1
By channel 1. The subsequent parameter
t-off
is used to specify the switch-off
threshold. The transmitter is switched off when the pressure on channel 1
exceeds the switch-off threshold.