50
GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
Parameters
LoG -3
Logarithmic representation of a partial measurement range (2.5 V/decade).
LoG +0
Logarithmic representation of a partial measurement range (2.5 V/decade).
LoG +3
Logarithmic representation of a partial measurement range (2.5 V/decade).
LoGC1
Logarithmic representation matched to the following sensor combination:
TTR on channel 1
PTR 225 on channel 2
LoGC2
Logarithmic representation matched to the following sensor combination:
CTR on channel 1
CTR on channel 2
This characteristic curve is only useful if the transmitters have different mea-
suring ranges. The total measuring range of the sensor combination is repre-
sented logarithmically in the range 0…10 V.
LoGC3
Logarithmic representation matched to the following sensor combination:
CTR on channel 1
CTR on channel 2
CTR on channel 3
This characteristic curve is only useful if the transmitters have different mea-
suring ranges. The total measuring range of the sensor combination is repre-
sented logarithmically in the range 0…10 V.
Note:
The three transmitters must be sorted with regard to their measuring
range (FS). The sort order may be increasing or decreasing.
Transmitter
Pressure (in mbar)
All types
p = 10^[U/(10/4) - 7]
Transmitter
Pressure (in mbar)
All types
p = 10^[U/(10/4) - 4]
Transmitter
Pressure (in mbar)
All types
p = 10^[U/(10/4) - 1]
Transmitter
Pressure (in mbar)
TTR + PTR 225
p = 10^[U/(10/12) - 9]