Computer interface
GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
77
6.3.26
SC1
Sensor 1 control. Transmitter 1 control. See Chapter 5.2.6 Transmitter activa-
tion (S-on),
44 and Chapter 5.2.8 Transmitter deactivation (S-oFF),
44.
S: SC1[,a,b,c.ccE±cc,d.ddE±dd]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c.ccE±cc,d.ddE±dd<CR><LF>
Note:
The commands SC2 and SC3 concern the transmitters 2 and 3, respec-
tively. The commands are analogous to the SC1 command.
6.3.27
SP1
Setpoint 1. Switching function 1. See Chapter 5.1 Switching function parame-
ters (PArA SP),
S: SP1[,a,b.bbbbE±bb,c.ccccE±cc]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b.bbbbE±bb,c.ccccE±cc<CR><LF>
Parameters
Significance
a
Transmitter activation
0 = Manual (default)
1 = Hot start
2 = By channel 1
3 = By channel 2
4 = By channel 3
b
Transmitter deactivation
0 = Manual (default)
1 = Self control
2 = By channel 1
3 = By channel 2
4 = By channel 3
c.ccE±cc
Activation value in current unit of measurement
d.ddE±dd
Deactivation value in current unit of measure-
ment
Parameters
Significance
a
Switching function assignment
0 = Channel 1
1 = Channel 2
2 = Channel 3