Operation
GA 09.033/6.02 – (2011-07) –
©
Oerlikon Leybold Vacuum
73
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
│
└─
Measurement range (F.S.)
Offset correction
Transmit:
OFS
[,x,x.xxxEsx] <CR>[<LF>]
│
│
│
└─
Offset
1)
│
[in current pressure unit]
│
(default = 0.000E0)
│
└─
Mode, x =
0 –> Off (default)
No offset value
needs to be entered.
1 –> On
If no offset value has been
entered, the previously
defined offset value is
taken
over.
2
–>
Auto
(offset
measurement)
No offset value needs to
be
entered.
3 -> Zero adjust CTR 100,
CTR 101. No offset value
needs to be entered.
1)
Values can be entered in any format. They are
internally converted into the floating point format.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x,sx.xxxxEsxx
<CR><LF>
│
│
│
└─
Offset
│
[in current pressure unit]
│
└─
Mode