Operation
58
GA 09.033/6.02 – (2011-07) –
©
Oerlikon Leybold Vacuum
A/D converter test 2
Test of channel 2 of the analog/digital converter (with a
reference voltage at the signal input of the SENSOR
connector (
→
24)).
The measurement value filter affects the ap-
plied voltage. If the signal input is open,
CENTER ONE displays a default value that
may easily fluctuate because of the high sensi-
tivity of the open measurement circuit.
Test sequence
z.B.:
Ö
Gauge identification voltage.
Ö
No gauge connected.