![Oerlikon CENTER ONE Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/oerlikon/center-one/center-one_operating-manual_3396572071.webp)
Operation
70
GA 09.033/6.02 – (2011-07) –
©
Oerlikon Leybold Vacuum
5.2.2 Parameter Mode
Degas
Transmit:
DGS
[,x] <CR>[<LF>]
│
└─
x = 0 –> off (default)
1 –> on (3 min.)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
│
└─
Degas status
Threshold value setting, allocation
Transmit:
SP1
[,x.xxEsx,x.xxEsx] <CR>[<LF>]
│
│
│
└─
Upper threshold
1)
│
[in current pressure
│
unit]
│
(default = depending
│
on gauge)
│
└─
Lower threshold
1)
[in current pressure unit]
(default = depending on
gauge)
1)
Values can be entered in any format. They are internally
converted into the floating point format.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x.xxxxEsxx,x.xxxxEsxx
<CR><LF>
│
│
│
└─
Upper threshold
│
[in current pressure unit]
│
└─
Lower threshold
[in current pressure unit]