Technical Data
12
GA 09.033/6.02 – (2011-07) –
©
Oerlikon Leybold Vacuum
Error signal (Error)
Number 1
Reaction time
≤
20 ms
Error signal relay
Contact type
floating normally open contact
Load max.
125 V
∼
, 60 W (ohmic)
110 V=, 2 A, 60 W (ohmic)
DANGER
For benchtop use, max. 30 V~
or 60 V= may be connected.
Service life
mechanic
electric
10
8
10
5
(at maximum load)
Connector
9-pin D-Sub, male
(pin assignment
→
26)
Contact positions
9
8
no error
9
8
error or no mains
power
Analog output
Number
1, recorder output
Voltage range
0 … +10 V
Deviation from
displayed value
±50 mV
Internal resistance
660
Ω
Relationship
measurement signal-
pressure
depending on gauge
(
→
[1] … [16])
Connector
9-pin D-Sub, male
(pin assignment
→
26)